Thermometer Industrial RTD assemblies TH11
Brand: Endress+HauserThermometer Industrial RTD assemblies TH11
Brand: Endress+Hauser
Extended Specs at a Glance
Max. immersion length on request:
up to 96" (2439 mm)
others on request
others on request
Response time:
63% rt = 2,0 s
Accuracy:
class A acc. to IEC 60751
class B acc. to IEC 60751
class B acc. to IEC 60751
Operating temperature range:
PT 100:
-200 °C ...600 °C
(-328 °F ...1.112 °F)
-200 °C ...600 °C
(-328 °F ...1.112 °F)
Max. process pressure (static):
at 20°C: 250 bar (3.626 psi)
depends on configuration
depends on configuration
Field of application
The TH11 is a modular thermometer. With different head transmitter it is a complete unit ready for use with enhanced measurement accuracy and reliability. It also enables very fast response times, due to direct contact measurement without thermowell. A variety of process connections and dimensions offer flexible application possibilities.
Specifications
Brand
Industry
- Chemical
- Water, Wastewater & Waste
- Power & Energy
- Mining, Minerals & Metals
- Pulp & Paper (incl. Printing)
- Other industries
Thermometer
Characteristic / Application
- US style
- modular temperature assembly
- threaded process connection
- direct immersion style
Thermowell / protection tube
- without (not intended to be used with thermowell)
Insert / probe
- mineral insulated (MI), flexible
- PTFE-insulated, rigid
Outer diameter protection tube / Insert
- 1/8'' (3,18 mm)
- 3/16''(4,76 mm)
- 1/4'' (6,35 mm)
- 3/8'' (9,53 mm)
Max. immersion length on request
- up to 96" (2439 mm)
- others on request
Material protection tube/ thermowell
- Sensor sheath
- 316/316L
Process connection
- male thread:
- G1/2"
- NPT1/2"
- compression fitting:
- NPT1/8"
- NPT1/4"
Tip shape
- straight
- reduced
Surface roughness Ra
- Not defined
Operating temperature range
- PT 100:
- -200 °C ...600 °C
- (-328 °F ...1.112 °F)
Max. process pressure (static)
- at 20°C: 250 bar (3.626 psi)
- depends on configuration
Accuracy
- class A acc. to IEC 60751
- class B acc. to IEC 60751
Response time
- 63% rt = 2,0 s
Integration head transmitter
- yes (4 … 20 mA; HART; PROFIBUS PA; FOUNDATION FIELDBUS)
Ex - approvals
- transmitter only
- FM IS
- CSA IS
- FM/CSA IS
Certification
- SIL (transmitter only)