Cerabar PMC71B

Brand: Endress+Hauser

Cerabar PMC71B

Brand: Endress+Hauser

Rust Automation & Controls, Endress+Hauser's exclusive USA representative in:
 CA CO HI MT   Parts of ID ND NV   UT WY

Extended Extended
Heartbeat Technology Heartbeat Technology

Specs at a Glance

Pressure measuring range:
100 mbar...40 bar
(1.5 psi...600 psi)
Process temperature:
-40°C...+150°C
(-40°F...+302°F)
Main wetted parts:
Ceraphire ceramic
FKM, EPDM, HNBR,
Perlast, Kalrez, Chemraz
Material process membrane:
Ceramic
Measuring cell:
100 mbar....40 bar
(1.5 psi...600 psi)
Accuracy:
Standard:
up to 0.05 %
Platinum:
up to 0.025 %

Field of application

This transmitter belongs to the new Cerabar generation. The device is robust and can be remotely controlled via app on a tablet using a secure Bluetooth connection. It allows condition based maintenance as well.The software is designed to simplify the handling. Intuitive wizards guide the user through setup, proof tests and verification of the device.Heartbeat Technology offers verification and monitoring functions to detect unwanted anomalies e.g. changes in loop characteristics.

Pressure measuring range: up to 40 bar (600 psi)

Fully vacuum resistant up to +150°C (302°F) process temperature

Accuracy: up to +/- 0.025%

Features

  • With Heartbeat Technology you are able to verify the health of the device while the process is running.
  • Simple indication of device status - display changes from green to red when diagnostic messages occur
  • Reduce systematic failures - error free SIL commissioning and instrument guided proof testing
  • Wireless control of the device in hard to reach process areas with the SmartBlue App

Specifications

Product Family

  • Cerabar

Continuous / Liquids

Characteristic / Application

  • Premium accuracy pressure transmitter with dry ceramic cell.
  • Intuitive and clear wizards for commissioning, proof tests and verification.

Specialities

  • Heartbeat Technology,
  • Bluetooth® Operation and maintenance SmartBlue App,
  • LED Module,
  • RFID TAG for easy identification,
  • Plug and play functionalities,
  • HistoROM

Supply / Communication

  • Ex d, Ex e, non-Ex::
  • 10.5...35V DC
  • Ex i: 10.5...30V DC

Accuracy

  • Standard:
  • up to 0.05 %
  • Platinum:
  • up to 0.025 %

Long term stability

  • 0.05 % of URL/ year
  • 0.08 % of URL/ 5 years
  • 0.10 % of URL/ 10 years

Ambient temperature

  • -40°C...+85°C
  • (-40°F...+185°F)

Process temperature

  • -40°C...+150°C
  • (-40°F...+302°F)

Process pressure / max. overpressure limit

  • 60 bar (900 psi)

Pressure measuring range

  • 100 mbar...40 bar
  • (1.5 psi...600 psi)

Main wetted parts

  • Ceraphire ceramic
  • FKM, EPDM, HNBR,
  • Perlast, Kalrez, Chemraz

Process connection

  • Thread:
  • G1/2...G2, R1/2, MNPT1/2...MNPT2, FNPT1/2
  • Flange:
  • DN25...DN80,
  • NPS 1"...4"

Max. measurement distance

  • 400 m (1312 ft) H2O

Communication

  • 4...20 mA HART
  • PROFIBUS PA
  • PROFINET with Ethernet-APL
  • Bluetooth® wireless
  • technology and app (optional)

Certificates / Approvals

  • ATEX, IEC Ex, CSA C/US, NEPSI

Safety approvals

  • SIL

Design approvals

  • EN 10204-3.1

Drinking water approvals

  • NSF, KTW

Metrological approvals and certificates

  • MID

Options

  • Touch control display

Application limits

  • Measuring cell: ceramics
  • If pressurized, possibly
  • use differential pressure
  • measurement with two
  • pressure transmitters
  • (electronic dp). Observe
  • ratio head pressure :
  • hydrostatic pressure

Pressure

Characteristic

  • Premium accuracy pressure transmitter with dry ceramic cell.
  • Intuitive and clear wizards for commissioning, proof tests and verification.

Supply voltage

  • Ex d, Ex e, non-Ex:
  • 10.5...35V DC
  • Ex i: 10.5...30V DC

Reference Accuracy

  • Standard:
  • up to 0.05 %
  • Platinum:
  • up to 0.025 %

Long term stability

  • 0.05 % of URL/ year
  • 0.08 % of URL/ 5 years
  • 0.10 % of URL/ 10 years

Process temperature

  • -40°C...+150°C
  • (-40°F...+302°F)

Ambient temperature

  • -40°C...+85°C
  • (-40°C...+185°F)

Measuring cell

  • 100 mbar....40 bar
  • (1.5 psi...600 psi)

Smallest calibratable span

  • 5 mbar (0.075 psi)

Vacuum resistance

  • 0 mbar abs.

Max. Turn down

  • 100:1

Max. overpressure limit

  • 60 bar (900 psi)

Process connection

  • Thread:
  • G1/2...G2, R1/2, MNPT1/2...MNPT2, FNPT1/2
  • Flange:
  • DN25...DN80,
  • NPS 1"...4"

Material process membrane

  • Ceramic

Material gasket

  • FKM, EPDM, HNBR, FFKM

Fill fluid

  • none, dry measuring cell

Material housing

  • Aluminium, 316L

Communication

  • 4...20 mA HART
  • PROFIBUS PA
  • PROFINET with Ethernet-APL
  • Bluetooth® wireless
  • technology and app (optional)

Certificates / Approvals

  • ATEX, CSA, IECEx, JPN Ex, INMETRO, NEPSI, UK Ex

Safety approvals

  • SIL

Design approvals

  • EN10204-3.1

Drinking water approvals

  • NSF, KTW

Specialities

  • Heartbeat Technology,
  • Bluetooth® Operation and maintenance SmartBlue App,
  • LED Module,
  • RFID TAG for easy identification,
  • Plug and play functionalities,
  • HistoROM
show more