Liquiline System CA80HA

Brand: Endress+Hauser

Liquiline System CA80HA

Brand: Endress+Hauser

Rust Automation & Controls, Endress+Hauser's exclusive USA representative in:
 CA CO HI MT   Parts of ID ND NV   UT WY

Extended Extended

Specs at a Glance

Measuring range:
0 to 80 mg/l CaCO3
0 to 80 mg/l with dilution function to maximum 16 to 1600 mg/l CaCO3
Process temperature:
4 to 40 °C (39 to 104 °F)
Process pressure:
Unpressurized

Field of application

Liquiline System CA80HA offers precise online analysis of water hardness in drinking water and process water. It helps you to optimize the control of water softening processes in drinking water applications and the optional dilution module allows water hardness measurements with concentration up to 1600 mg/l. The analyzer's automatic calibration and cleaning and its low reagent consumption reduce operating costs while advanced diagnostics with remote access ensure consistent process safety.

Liquiline System CA80HA supports you in monitoring total water hardness:

  • To classify water hardness in drinking water
  • To ensure process water quality in production plants
  • For monitoring process water, e.g. in pure and ultrapure water or desalination plants, requiring a lower limit of detection, please refer to the low range analyzer Liquiline System CA82HA.

Features

  • The water hardness analyzer uses the colorimetric phthalein purple method, ensuring direct comparability to lab inputs
  • Fast commissioning with plug & play thanks to Memosens technology and user-friendly operation
  • Reduces operating costs thanks to automatic calibration and cleaning, low reagent consumption and long reagent lifetime with optional cooling module
  • Easy upgrade of functionality to a complete measuring station - by adding modules and connecting Memosens sensors.
  • Advanced diagnostics with remote access for increased process safety
  • Easy, tool-free maintenance

Specifications

Industry

  • Water, Wastewater & Waste
  • Power & Energy
  • Oil & Gas
  • Pulp & Paper (incl. Printing)

Product Family

  • Liquiline

Analyser

Characteristic

  • Process analyzer for hardness in aqueous solutions
  • Comply with standard colorimetric measuring principle - phthalein purple method

Size

  • Housing (open version):
  • 793 x 530 x 417 mm
  • 31.22 x 20.87 x 16.42 in
  • Housing (closed version):
  • 793 x 530 x 463 mm
  • 31.22 x 20.87 x 18.23 in
  • Housing with base:
  • 1723 x 530 x 463 mm
  • 67.83 x 20.87 x 18.23 in

Design

  • Open design, cabinet and stand housing
  • High-Performance plastic ASA-PC, additional stand coated steel

Process temperature

  • 4 to 40 °C (39 to 104 °F)

Ambient temperature

  • 5 to 40 °C (41 to 104 °F)

Process pressure

  • Unpressurized

Sample flow rate

  • Min. 5 ml/min (0.17 fl.oz/min)

Consistency of the sample

  • Low solid content (< 50NTU), aqueous homogeneous

Application

  • Total hardness monitoring in reverse osmosis systems inlets, after ion exchange systems
  • Monitoring of water preparation in processes and drinking water

Power supply

  • 100 ... 120 VAC / 200 ... 240 VAC ± 10%
  • 24 VDC ± 10%
  • 50 ± 1 oder 60 ± 1,2 Hz

Output / communication

  • 2x 0/4 ... 20 mA
  • Webserver, Modbus, Ethernet/IP, Profibus DP

Input

  • 1 or 2 measuring channel
  • 1 ... 4 digital sensor inputs for sensors with Memosens protocol (optional)

Measuring range

  • 0 to 80 mg/l CaCO3
  • 0 to 80 mg/l with dilution function to maximum 16 to 1600 mg/l CaCO3

Consumables

  • Reagents and standard solutions CY80HA as well as cleaner CY800 are necessary for the operation
  • Regular maintenance is done with the parts of the maintenance kit CAV800
show more