Extended Specs at a Glance
Measuring range:
100 µS/cm to 2000 mS/cm
Cell constant k: 6.3 1/cm
Cell constant k: 6.3 1/cm
Process pressure:
13 bar abs up to 90 °C (188.5 psi abs up to 194 °F)
9 bar at 125 °C (130.5 psi abs at 257 °F)
9 bar at 125 °C (130.5 psi abs at 257 °F)
Process temperature:
-10 °C to 125 °C (14 °F to 257 °F)
Sterilization: 150 °C at 6 bar abs for max. 60 min
(302 °F at 87 psi abs for max. 60 min)
Sterilization: 150 °C at 6 bar abs for max. 60 min
(302 °F at 87 psi abs for max. 60 min)
Field of application
Conductivity sensorRequest order codesThe Indumax CLS54D inductive conductivity sensor protects all of your products and processes that must meet the strictest hygienic and sterile demands. Its outstanding performance combined with a unique hygienic design ensures highest safety and quality of your products. Thanks to Memosens digital technology, Indumax CLS54D combines maximum process and data integrity with simple operation. It resists corrosion and moisture, enables lab calibration, and facilitates predictive maintenance.
Specifications
Brand
Industry
- Food & Beverage
- Other industries
- Life Science
Product Family
- Indumax
Conductivity
Application
- Food, beverage and pharmaceutical industries, biotechnology, dairies, breweries
Characteristic
- Specifically intended for use in hygienic applications and especially suitable for following applications:
- Phase separation, control of CIP, concentration control, product monitoring, leakage monitoring
Measuring range
- 100 µS/cm to 2000 mS/cm
- Cell constant k: 6.3 1/cm
Measuring principle
- Inductive conductivity measurement
Design
- Hygienic design with food-grade virgin PEEK body in jointless and creviceless design including all hygienic certifications
Material
- In contact with medium: virgin PEEK
Process temperature
- -10 °C to 125 °C (14 °F to 257 °F)
- Sterilization: 150 °C at 6 bar abs for max. 60 min
- (302 °F at 87 psi abs for max. 60 min)
Process pressure
- 13 bar abs up to 90 °C (188.5 psi abs up to 194 °F)
- 9 bar at 125 °C (130.5 psi abs at 257 °F)
Temperature sensor
- Pt1000 (Class A according to IEC 60751)
Connection
- Dairy pipe, Aseptic-fitting, Clamp ISO 2852, SMS-2", Varivent N DN 40-125, Neumo Biocontrol D50
Ingress protection
- IP68 / NEMA type 6P
Additional certifications
- FDA, EHEDG, 3-A
- Biological reactivity USP class VI part <87> and <88>88>87>