Conductivity sensor Memosens CLS15E
Brand: Endress+HauserConductivity sensor Memosens CLS15E
Brand: Endress+Hauser
Extended Specs at a Glance
Measurement range:
k=0,01: 0.04 to 20 µS/cm
k=0,1: 0.10 to 200 µS/cm
k=0,1: 0.10 to 200 µS/cm
Process temperature:
Threaded with fixed cable:
-20 to 100 °C (-4 to 212 °F)
Threaded with plug-in head:
-20 to 120 °C (-4 to 248 °F)
Sterilization: max. 140 °C (284 °F) for 30 minutes
-20 to 100 °C (-4 to 212 °F)
Threaded with plug-in head:
-20 to 120 °C (-4 to 248 °F)
Sterilization: max. 140 °C (284 °F) for 30 minutes
Process pressure:
13 bar at 20 °C (188 psi at 68 °F) absolute
1 bar at 120 °C (14 psi at 248 °F) absolute
1 bar at 120 °C (14 psi at 248 °F) absolute
Field of application
The Memosens CLS15E is the ideal choice for measuring conductivity in environments with low measuring ranges, such as boiler feed water and chip cleaning. This contact-based conductivity sensor ensures reliable and accurate performance, even in hazardous environments. With its Memosens 2.0 digital technology, the CLS15E has an extended memory for calibration, setting and process data, enabling predictive maintenance and providing an excellent basis for IIoT services.
Features
- Designed for durability, the sensor requires minimal maintenance and is characterized by a long service life.
- Its sterilizable and autoclavable properties allow installation in sterile environments within plants.
- With its electrode geometry, the Memosens CLS15E provides accurate readings even at low conductivities, ensuring reliability.
- A certificate ensuring the specific cell constant allows for exact fine-tuning of the measuring point, ensuring precision in adjustment.
- IIoT-enabled with Memosens 2.0, which offers extended storage of calibration and process data and enables improved trend analysis, predictive maintenance and advanced IIoT services.
Specifications
Brand
Product Family
- Memosens
Conductivity
Application
- Measurement in pure and ultrapure water range
- Monitoring of ion exchangers
- Reverse osmosis
- Distillation and chip cleaning
Characteristic
- Digital 2-electrode conductivity sensor
Measurement range
- k=0,01: 0.04 to 20 µS/cm
- k=0,1: 0.10 to 200 µS/cm
Measuring principle
- Conductive conductivity cell with electropolished stainless steel electrodes
Design
- 2-electrode conductivity cell with coaxially arranged electrodes, electropolished
Material
- Electrode: stainless steel 1.4435
- Electrode shaft: PES
Dimension
- Electrode diameter: 16 mm (0.63 inch)
- Electrode length: appr. 55 mm (2.17 inch)
Process temperature
- Threaded with fixed cable:
- -20 to 100 °C (-4 to 212 °F)
- Threaded with plug-in head:
- -20 to 120 °C (-4 to 248 °F)
- Sterilization: max. 140 °C (284 °F) for 30 minutes
Process pressure
- 13 bar at 20 °C (188 psi at 68 °F) absolute
- 1 bar at 120 °C (14 psi at 248 °F) absolute
Temperature sensor
- Pt1000
Ex certification
- ATEX, NEPSI, CSA, IECEx, INMETRO, EAC Ex
Connection
- Process: 1/2" and 3/4" NPT, Clamp 1 1/2"
- Sensor connection: Inductive, digital connection head with Memosens 2.0 technology
Ingress protection
- IP68
Additional certifications
- Calibration certification of the cell constante and temperature, material certification 3.1