Liquiline System CA76NA

Brand: Endress+Hauser

Liquiline System CA76NA

Brand: Endress+Hauser

Rust Automation & Controls, Endress+Hauser's exclusive USA representative in:
 CA CO HI MT   Parts of ID ND NV   UT WY

Specs at a Glance

Process temperature:
10 to 40 °C
(50 to 104 °F)
Process pressure:
1 to 5 bar (abs)
14.5 to 72.5 psi
Measuring range:
0.1 to 9999 µg/l (ppb) Na

Field of application

Protect your water steam cycles and ion exchangers with the CA76NA sodium analyzer! An increase in sodium levels in high-purity or boiler feedwater indicates the presence of unwanted dissolved impurities. This can cause serious damage and consequently expensive repairs. CA76NA provides highly accurate online monitoring of water quality and enables fast detection of critical events. The sodium analyzer supports your preventive maintenance strategy to keep your power plant running.

Features

  • Indication of the purity of water and steam used in boilers and turbines
  • Provision of early warning for condenser leakages or malfunction of ion exchangers
  • Control of the incoming water quality in power plants that use desalination plants as water source

Specifications

Industry

  • Power & Energy

Analyser

Characteristic

  • Analyser for sodium in boiler feedwater, steam, condensate and ion exchangers

Size

  • 850 x 450 x 250 mm
  • 33.5 x 17.7 x 9.8 inch

Design

  • Panel stainless steel 1.4301

Process temperature

  • 10 to 40 °C
  • (50 to 104 °F)

Ambient temperature

  • 5 to 40 °C
  • (41 to 104 °F)

Process pressure

  • 1 to 5 bar (abs)
  • 14.5 to 72.5 psi

Sample flow rate

  • 10 to 15 l/h
  • 2.64 to 3.96 gal/h

Application

  • Online sodium monitoring in boiler feed water, water steam cycles, condensate and ion exchangers

Power supply

  • 100 to 240 V AC, 50/60 Hz

Output / communication

  • 0/4 to 20 mA

Measuring range

  • 0.1 to 9999 µg/l (ppb) Na

Consumables

  • CA76NA sodium electrode,
  • CA76NA pH combination electrode
  • CA76NA standard solution 5100 ppb Na
show more