S030 In-Line Sensor-Fitting with Paddle Wheel
Brand: Burkert Fluid Control SystemsThe sensor-fitting S030 has a built-in paddle wheel to measure the flow rate. When liquid flows through the pipe, the paddle wheel is set in rotation producing pulses which frequency is proportional to the flow rate.
The Burkert ”INLINE Quarter-turn” technology is a construction ensuring a leakage free operation. The paddle wheel rotation (permanent magnets included in the wheels) is detected contactless through the sensor-fitting wall. The transmitter can be snapped-on or removed without opening the pipe or interrupting the process.
Features
- DN06 to DN65
- Closed pipe system, sensor inside fitting
- Quarter-turn technology
- Transmitter available for:
- Indication, Monitoring, Transmitting
- On/Off control, Batch control
Specifications
Wetted Material
- Sensor Armature: 316L Stainless Steel 1.4404, Brass, PVC, PP or PVDF
- Axis: Ceramics (Al2O3)
- Polypropylene (PP)
- PVC
- PVDF
- Bearing: Ceramics (Al2O3), Iglidur®
- Paddle Wheel: PVDF, Stainless Steel (316L ‑ 1.4404), PP on request
- Sensor-fitting Body: 316L Stainless Steel (1.4404) , Brass, PVC, PP, PVDF
Seal Material (Wetted)
- EPDM, FKM
Screw Material (Non-Wetted)
- Stainless Steel (316L ‑ 1.4404)
Pipe Diameter
- DN 06 to DN 65
Measuring Range
- 0.5 to 1,200 L/minute (0.13 to 317 gpm)
Measuring Deviation
- Teach-In (Via A Remote Transmitter): ± 1 % of the Measured Value
- Standard K factor: ± 2.5 % of the Measured Value
Linearity
- ± 0.5 % of Full Scale
Repeatability
- ± 0.4 % of the Measured Value
Media
- Clean, Netural or Slightly Aggressive, Solid-free Liquids
Viscosity
- Max. 300 cSt
Medium Temperature
- Sensor-fitting in PP: 0° to 80° C (32° to 176° F)
- Sensor-fitting in PVC: 0° to 50° C (32° to 122° F)
- Sensor-fitting in Stainless Steel, Brass or PVDF: -15° to 100° C (5° to 212° F)
Fluid Pressure
- Max PN 10 to PN 40
Ambient Temperature
- Operating & Storage (For Sensor-fitting in PP): -15° to 80° C (5° to 176° F)
- Operating & Storage (For Sensor-fitting in PVC): -15° to 60° C (5° to 140° F)
- Operating & Storage (For Sensor-fitting in Stainless Steel, Brass or PVDF): -15° to 100° C (5° to 212° F)
Flow Rate
- 0.3 to 10 m/sec
Approvals & Certifications
- CE
- FDA
- PED 2014/68/EU