Memosens CPS71E

Brand: Endress+Hauser

Memosens CPS71E

Brand: Endress+Hauser

Rust Automation & Controls, Endress+Hauser's exclusive USA representative in:
 CA CO HI MT   Parts of ID ND NV   UT WY

Extended Extended

Specs at a Glance

Process temperature:
1 to 140 °C (32 to 284 °F)
Measurement range:
Application B:
• pH: 0 to 14
Application H:
• pH: 0 to 12
Process pressure:
0,8 to 14 bar (11,6 to 203 psi) (absolute)

Field of application

The Memosens CPS71E is designed for demanding processes. Its contamination-resistant reference and moisture resistance ensure reliable measurements in heavily contaminated, aggressive media such as strong acids and bases. With Memosens 2.0 digital technology, the CPS71E increases the storage capacity for calibration and process data and supports predictive maintenance. Laboratory calibration and quick on-site sensor replacement maximize process uptime.

Features

  • The special ion trap of the electrode prevents poisoning of the junction and the reference, offers resistance to strong acids and bases and ensures a longer service life of the sensor.
  • The optional pressurized reference ensures reliable measurements in blocking media.
  • Optional upside-down mounting for flexible installation.
  • Inductive cable connection and non-contact signal transmission improve process integrity by preventing problems with moisture or corrosion.
  • On-site sensor replacement minimizes process downtime and costs.
  • Memosens 2.0 technology enables trend identification, predictive maintenance and advanced IIoT services with extended data storage for calibration and process information.

Specifications

Product Family

  • Memosens

pH

Application

  • Process technology and monitoring of processes with:
  • • Rapidly changing pH values
  • • High proportion of electrode poisons such as H2S

Characteristic

  • Digital pH electrode for chemical process with an ion trap for poison-resistant reference

Measurement range

  • Application B:
  • • pH: 0 to 14
  • Application H:
  • • pH: 0 to 12

Measuring principle

  • Gel compact electrode with ceramic junction and ion trap

Design

  • All shaft lengths with temperature sensor
  • Advanced gel technology

Material

  • Sensor shaft: Glass to suit process
  • pH membrane glass: Type B, Type N
  • Metal lead: Ag/AgCl
  • Open aperture: Ceramic junction, zirconium dioxide
  • O-ring: FKM
  • Process coupling: PPS fiber-glass reinforced
  • Nameplate: Ceramic metal oxide

Dimension

  • Diameter: 12 mm (0.47 inch)
  • Shaft lengths: 120, 225, 360 and 425 mm
  • (4.72, 8.86, 14.2 and 16.7 inch)

Process temperature

  • 1 to 140 °C (32 to 284 °F)

Process pressure

  • 0,8 to 14 bar (11,6 to 203 psi) (absolute)

Temperature sensor

  • NTC 30k

Ex certification

  • With ATEX, IECEx, CSA C/US, NEPSI, Japan Ex and INMETRO approvals for use in
  • hazardous areas Zone 0, Zone 1 and Zone 2.

Connection

  • Inductive, contactless connection head with Memosens 2.0 technology

Ingress protection

  • IP68

Additional certifications

  • Additional certifications

Process Technology and Monitoring of Processes with:
  • Rapidly Changing pH Values
  • High Proportion of Electrode Poisons Such As H2S
show more