MS05 Turbidity Sensor Cube

Brand: Burkert Fluid Control Systems

The device is a turbidity measurement sensor. It is used within the Online Analysis System Type 8905 by being plugged into a spare fluidic backplane slot. 

The device contains an optical sensor following DIN EN ISO 7027 or EPA method 180.1. The turbidity of water needs to be analyzed continuously as indicator for unwanted undissolved content in the water. 

Measurement prior and post filtration indicates the effect of filtration and may help to optimize the filter backwash process. In the best case it can lead to water and energy savings. The electrical and fluidic connections are made via the connection panel of the system. 

The sensor cube is communicating via buS, so the recognition at the Online Analysis System is fully automatic. When plugged into a system you will find the sensor in the list of buS members for further customized adjustments.

Features

  • Fully compatible with buS systems and a wide range of further analysis sensor cubes
  • Optical sensor according to DIN EN ISO 7027 or EPA method 180.1
  • Modular sensor cube for hot swap (exchange during operation)
  • Minimal sample water flow needed

Specifications

Seal Material

  • EPDM

Housing Material

  • PPE + PS

Materials of Construction

  • Cuvette: Glass
  • Lever: Zamak, Painted
  • Valve: Silicone

Measuring Range

  • 0 to 40 FNU

Measuring Range Resolution

  • ± 0.0006 FNU

Measuring Deviation

  • ± 0.02 FNU or 2 % of Reading, whichever is greater

Repeatability

  • ± 0.02 FNU or 2 % of Reading, whichever is greater

Linearity

  • ± 0.5 % of Full Scale

Operating Voltage

  • 24 Vdc

Power Consumption

  • 0.8 VA

Ambient Temperature

  • Operation: 3° to 40° C (37° to 104° F)
  • Storage & Transport: -10° to 60° C (14° to 140° F)

Humidity

  • < 90 % RH, without Condensation

Pressure of the Fluid Sample

  • psi: 43.51 (kPa: 300, bar: 3)

Temperature of the Fluid Sample

  • 3° to 40° C (37° to 104° F)

Electrical Connection

  • Spring Contacts

Media

  • Water without particles: drinking water, industrial water

Environmental Protection

  • IP20
  • IP65

Approvals & Certifications

  • CE
  • EU Directives
  • NAMUR NE 107 (External Communication by Status LED)
  • UL (C,US)

Process Connection

  • Via pinch valve in the fluidic backplane of Type 8905
show more