Lean Specs at a Glance
Measuring cell:
400 mbar...400 bar
(6psi...6,000psi)
(6psi...6,000psi)
Accuracy:
Standard:
up to 0.075 %
Platinum
up to 0.055 %
up to 0.075 %
Platinum
up to 0.055 %
Pressure measuring range:
400 mbar... 400 bar
(6psi...6,000psi)
(6psi...6,000psi)
Material process membrane:
316L, AlloyC,
Tantal, Monel,
PTFE
Tantal, Monel,
PTFE
Process temperature:
Standard:
-40°C…+125°C
(-40°F…+257°F)
Diaphragm seal:
-40°C...+400°C
(-40°F...+752°F)
-40°C…+125°C
(-40°F…+257°F)
Diaphragm seal:
-40°C...+400°C
(-40°F...+752°F)
Main wetted parts:
316L, AlloyC,
Tantal, Monel,
PTFE
Tantal, Monel,
PTFE
Field of application
The Cerabar PMP51B is a pressure transmitter with metal membrane for highly accurate measurement of liquids and gases. It offers easy local or remote operation and condition-based maintenance which promotes better management of scheduled plant shutdowns.
Specifications
Brand
Product Family
- Cerabar
Continuous / Liquids
Characteristic / Application
- Pressure transmitter with metal membrane for accurate measurement in liquids or gases.
- Available with diaphragm seals.
- Intuitive and clear wizard for commissioning
Specialities
- Bluetooth® Operation and maintenance SmartBlue App,
- RFID TAG for easy identification,
- Plug and play functionalities
Supply / Communication
- Ex d, Ex e, non-Ex:
- 10.5...35V DC
- Ex i: 10.5...30V DC
Accuracy
- Standard:
- up to 0.075 %
- Platinum
- up to 0.055 %
Long term stability
- 0.1 % of URL/ year
- 0.2 % of URL/ 5 years
- 0.25 % of URL/ 10 years
Ambient temperature
- -40°C...+85°C
- (-40°F...+185°F)
Process temperature
- Standard:
- -40°C…+125°C
- (-40°F…+257°F)
- Diaphragm seal:
- -40°C...+400°C
- (-40°F...+752°F)
Process pressure / max. overpressure limit
- 700 bar (10,500 psi)
Pressure measuring range
- 400 mbar... 400 bar
- (6psi...6,000psi)
Main wetted parts
- 316L, AlloyC,
- Tantal, Monel,
- PTFE
Process connection
- Thread:
- G1/2...G2, R1/2, MNPT3/4...MNPT2,
- FNPT1/2
- Flange:
- DN25...DN80,
- NPS 1"...3"
Max. measurement distance
- 4000 m (13,123 ft) H2O
Communication
- 4...20 mA,
- 4...20 mA HART
- PROFINET with Ethernet-APL
- Bluetooth® wireless
- technology and app (optional)
Certificates / Approvals
- ATEX, IEC Ex, CSA C/US, NEPSI
Safety approvals
- SIL
Design approvals
- EN 10204-3.1
- NACE MR0175,
- NACE MR0103,
- AD2000
Hygienic approvals
- Hygienic approvals
Marine approval
- Marine approval
Drinking water approvals
- NSF
Options
- Diaphragm seal
Application limits
- Measuring cell: Metal welded
- If pressurized, possibly use differential pressure measurement with two pressure transmitters (electronic dp). Observe ratio head pressure: hydrostatic pressure
Pressure
Characteristic
- Pressure transmitter with metal membrane for accurate measurement in liquids or gases.
- Available with diaphragm seals.
- Intuitive and clear wizard for commissioning
Supply voltage
- Ex d, Ex e, non-Ex:
- 10.5...35V DC
- Ex i: 10.5...30V DC
Reference Accuracy
- Standard:
- up to 0.075 %
- Platinum:
- up to 0.055 %
Long term stability
- 0.1% of URL/year
- 0.2% of URL/ 5 years
- 0.25% of URL/ 10 years
Process temperature
- Standard:
- -40°C…+125°C
- (-40°F…+257°F)
- Diaphragm seal:
- -70°C...+400°C
- (-x°F...+752°F)
Ambient temperature
- -40°C...+85°C
- (-40°F...+185°F)
Measuring cell
- 400 mbar...400 bar
- (6psi...6,000psi)
Smallest calibratable span
- 20 mbar (0.3 psi)
Vacuum resistance
- 50 mbar (0.725 psi)
Max. Turn down
- 100:1
Max. overpressure limit
- 2,800 bar (40,600 psi)
Process connection
- Thread:
- G1/2...G2, R1/2, MNPT3/4...MNPT2,
- FNPT1/2
- Flange:
- DN25...DN80,
- NPS 1"...3"
Process connection hygienic
- Process connection hygienic
Material process membrane
- 316L, AlloyC,
- Tantal, Monel,
- PTFE
Material gasket
- PTFE
Fill fluid
- Silicone oil, Inert oil,
- Vegetable oil,
- High/ low temperature oil
Material housing
- Aluminium
Communication
- 4...20 mA,
- 4...20 mA HART
- PROFINET with Ethernet-APL
- Bluetooth® wireless
- technology and app (optional)
Certificates / Approvals
- ATEX, IEC Ex, CSA C/US, NEPSI
Safety approvals
- SIL
Design approvals
- EN10204-3.1,
- NACE MR0175,
- NACE MR0103
Hygienic approvals
- Hygienic approvals
Marine approvals
- Marine approvals
Drinking water approvals
- NSF
Specialities
- Bluetooth® Operation and maintenance SmartBlue App,
- RFID TAG for easy identification,
- Plug and play functionalities