Fluke Calibration 9171 Field Metrology Well

Brand: Fluke Corporation

Dry-wells are typically calibrated by inserting a calibrated PRT into one of the wells and making adjustments to the calibrator’s internal control sensor based on the readings from the PRT. 

This has limited value because the unique characteristics of the reference PRT, which essentially become “calibrated into” the calibrator, are often quite different from the thermometers tested by the calibrator. 

This is complicated by the presence of significant thermal gradients in the block and inadequate sensor immersion into blocks that are simply too short. 

Metrology Wells are different. Temperature gradients, loading effects, and hysteresis have been minimized to make the calibration of the display much more meaningful. 

We use only traceable, accredited PRTs to calibrate Metrology Wells and our proprietary electronics consistently demonstrate repeatable accuracy more than ten times better than our specs, which range from ±0.1 °C at the most commonly used temperatures to ±0.25 °C at 661 °C. 

 For even better accuracy, Metrology Wells may be ordered with built-in electronics for reading external PRTs with ITS-90 characterizations.
Features
  • Best-performing industrial heat sources (accuracy, stability, uniformity) in the world
  • Immersion depth to 203 mm (8 in)
  • Optional ITS-90 reference input reads PRTs to ±0.006 °C
  • Temperature range from –45 °C to 700 °C

Specifications

Range at 23° C Ambient

  • -30° C to 155° C
  • -22° F to 311° F

Display Accuracy

  • ± 0.1° C Full Range

Stability

  • ± 0.005° C Full Range

Axial Uniformity

  • ± 0.025° C at -30° C
  • ± 0.02° C at 0° C
  • ± 0.07° C at 155° C

Radial Uniformity

  • ± 0.01° C Full Range

Loading Effect

  • ± 0.005° C at -30° C
  • ± 0.005° C at 0° C
  • ± 0.01° C at 155° C

Hysteresis

  • 0.025° C

Resolution

  • 0.001° C

Computer Interface

  • RS-232 included

Temperature Range

  • -200° C to 962° C
  • -328° F to 1764° F

Resistance Range

  • 0 Ω to 400 Ω, auto-ranging

Resistance Accuracy

  • 0 Ω to 20 Ω: 0.0005 W
  • 20 Ω to 400 Ω: 25 ppm

show more