Memosens CLS82E
Brand: Endress+HauserMemosens CLS82E
Brand: Endress+Hauser
Extended Specs at a Glance
Process pressure:
17 bar abs at 20 °C (247 psi at 68 °F)
9 bar abs at 120 °C (131 psi at 248 °F)
9 bar abs at 120 °C (131 psi at 248 °F)
Process temperature:
-5 to 120 °C (23 to 248 °F)
Sterilization: max. 140 °C at 6 bar for max. 45 min
(Max. 284 °F at 87 psi for max. 45 min)
Sterilization: max. 140 °C at 6 bar for max. 45 min
(Max. 284 °F at 87 psi for max. 45 min)
Measurement range:
1 µS/cm to 500 mS/cm
Field of application
The Memosens CLS82E is a high-end, hygienic sensor for applications with widely varying conductivity values. It enables reliable monitoring of core processes as well as final rinse with only one sensor. FDA compliance and a small, compact design ensure the perfect fit to your processes. With Memosens 2.0 digital technology, CLS82E allows for extended storage of process and sensor data facilitating lab calibration and predictive maintenance and providing the perfect basis for enhanced IIoT services.
Specifications
Brand
Product Family
- Memosens
Conductivity
Application
- Phase seperation, chromotography, fermentation
- CIP monitoring in small pipes
- Ultrafiltration
- Final rinse
Characteristic
- Conductivity sensor with a broad measurement range for all hygienic applications with advanced requirements to process safety
Measurement range
- 1 µS/cm to 500 mS/cm
Measuring principle
- Hygienic 4-electrode conductivity sensor
Design
- Certified hygienic design with FDA compliant materials, 3-A and EHEDG certified, available as 120 mm sensor with PG13.5 and in many other standard process connections
Material
- Process connection: stainless steel 1.4435 (316L)
- Sensor element: platinum and ceramic
Dimension
- Sensor diameter: 12 mm (0.47 inch)
- Electrode length: depending on process connection
Process temperature
- -5 to 120 °C (23 to 248 °F)
- Sterilization: max. 140 °C at 6 bar for max. 45 min
- (Max. 284 °F at 87 psi for max. 45 min)
Process pressure
- 17 bar abs at 20 °C (247 psi at 68 °F)
- 9 bar abs at 120 °C (131 psi at 248 °F)
Temperature sensor
- Pt1000
Ex certification
- ATEX, NEPSI, CSA, IECE, EAC Ex, INMETRO
Connection
- Clamp 1,5", Clamp 2", Varivent F DN25,Varivent N DN40-DN125, BioControl DN25, PG13,5, DN25 standard/40mm, DN25 B.Braun port/65mm, Threat G1", Thread NPT 1"
- Sensor connection: Inductive, digital connection head with Memosens 2.0 technology
Ingress protection
- IP 68 / NEMA Type 6P
Additional certifications
- Material certification 3.1
- 3-A and EHEDG certified, hygienic design
- ASME BPE
- Conformity to cGMP derived requirements
- FDA approved
- Regulation (EC) No. 1935/2004
- China Food