Cerabar PMP71B

Brand: Endress+Hauser

Cerabar PMP71B

Brand: Endress+Hauser

Rust Automation & Controls, Endress+Hauser's exclusive USA representative in:
 CA CO HI MT   Parts of ID ND NV   UT WY

Extended Extended
Heartbeat Technology Heartbeat Technology

Specs at a Glance

Main wetted parts:
316L, AlloyC,
Tantal, Monel,
PTFE, Gold
Accuracy:
Standard:
up to 0.05 %
Platinum:
up to 0.025 %
Pressure measuring range:
400 mbar...700 bar
(1.5 psi...10,500 psi)
Process temperature:
Standard:
-40°C…+125°C
(-40°F…+257°F)
Diaphragm seal:
-40°C...+400°C
(-40°F...+752°F)
Measuring cell:
400 mbar...700 bar
(6 psi...10,500 psi)
Material process membrane:
316L, AlloyC,
Tantal, Monel,
PTFE,
Gold

Field of application

This transmitter belongs to the new Cerabar generation. The transmitter is robust and can be remotely controlled via app on a mobile device using a secure Bluetooth connection.The software is designed to simplify the handling. Intuitive wizards guide the user through setup, proof tests and verification of the device.Heartbeat Technology offers verification and monitoring functions to detect unwanted anomalies e.g. changes in loop characteristics.

Pressure measuring range: up to 700 bar (10500 psi)

Process temperatures up to 400°C (752 °F) with diaphragm seal

Accuracy: up to +/- 0.025%

Features

  • With Heartbeat Technology you are able to verify the health of the device while the process is running.
  • Simple indication of device status - display changes from green to red when diagnostic messages occur
  • Reduce systematic failures - error free SIL commissioning and instrument guided proof testing
  • Wireless control of the device in the process area with the SmartBlue App

Specifications

Product Family

  • Cerabar

Continuous / Liquids

Characteristic / Application

  • Pressure transmitter with metallic process diaphragm.
  • Available with diaphragm seals.
  • Device health verification without process interruptions.
  • Intuitive and clear wizards for commissioning, proof tests and verification.

Specialities

  • Heartbeat Technology,
  • Bluetooth® Operation and maintenance SmartBlue App,
  • LED Module,
  • RFID TAG for easy identification,
  • Plug and play functionalities,
  • HistoROM

Supply / Communication

  • Ex d, Ex e, non-Ex:
  • 10.5...35V DC
  • Ex i: 10.5...30V DC

Accuracy

  • Standard:
  • up to 0.05 %
  • Platinum:
  • up to 0.025 %

Long term stability

  • 0.05 % of URL/ year
  • 0.07 % of URL/ 5 years
  • 0.10 % of URL/ 10 years

Ambient temperature

  • -60°C...+80°C
  • (-76°F...+185°F)

Process temperature

  • Standard:
  • -40°C…+125°C
  • (-40°F…+257°F)
  • Diaphragm seal:
  • -40°C...+400°C
  • (-40°F...+752°F)

Process pressure / max. overpressure limit

  • 700 bar
  • (10,500 psi)

Pressure measuring range

  • 400 mbar...700 bar
  • (1.5 psi...10,500 psi)

Main wetted parts

  • 316L, AlloyC,
  • Tantal, Monel,
  • PTFE, Gold

Process connection

  • Thread:
  • G1/2...G2, R1/2, MNPT3/4...MNPT2,
  • FNPT1/2, R1/2
  • Flange:
  • DN25...DN100,
  • NPS 1"...4"
  • Pancake: 2"...4", DN50...DN100

Max. measurement distance

  • 7000 m (22,966 ft) H2O

Communication

  • 4...20 mA HART
  • PROFIBUS PA
  • PROFINET with Ethernet-APL
  • Bluetooth® wireless
  • technology and app (optional)

Certificates / Approvals

  • ATEX, IEC Ex, CSA C/US, NEPSI

Safety approvals

  • SIL

Design approvals

  • EN 10204-3.1
  • NACE MR0175,
  • NACE MR0103

Marine approval

  • Marine approval

Drinking water approvals

  • NSF

Metrological approvals and certificates

  • MID

Options

  • Touch control display
  • Diaphragm seal

Application limits

  • Measuring cell: Metal
  • welded
  • If pressurized, possibly
  • use differential pressure
  • measurement with two
  • pressure transmitters
  • (electronic dp). Observe
  • ratio head pressure:
  • hydrostatic pressure

Pressure

Characteristic

  • Pressure transmitter with metallic process diaphragm.
  • Available with diaphragm seals.
  • Device health verification without process interruptions.
  • Intuitive and clear wizards for commissioning, proof tests and verification.

Supply voltage

  • 4...20 mA HART
  • Ex d, Ex e, non-Ex:
  • 10.5...35V DC :
  • Ex i: 10.5...30V DC

Reference Accuracy

  • Standard:
  • up to 0.05 %
  • Platinum:
  • up to 0.025 %

Long term stability

  • 0.05 % of URL/ year
  • 0.07 % of URL/ 5 years
  • 0.10 % of URL/ 10 years

Process temperature

  • Standard:
  • -40°C…+125°C
  • (-40°F…+257°F)
  • Diaphragm seal:
  • -70°C...+400°C
  • (-94°F...+752°F)

Ambient temperature

  • -60°C...+80°C
  • (-76°F...+185°F)

Measuring cell

  • 400 mbar...700 bar
  • (6 psi...10,500 psi)

Smallest calibratable span

  • 5 mbar (0.075 psi)

Vacuum resistance

  • 5 mbar (0.725 psi)

Max. Turn down

  • 100:1

Max. overpressure limit

  • 1,050 bar (15,750 psi)

Process connection

  • Thread:
  • G1/2...G2, R1/2, MNPT3/4...MNPT2,
  • FNPT1/2, R1/2
  • Flange:
  • DN25...DN100,
  • NPS 1"...4"
  • Pancake: 2"...4", DN50...DN100

Material process membrane

  • 316L, AlloyC,
  • Tantal, Monel,
  • PTFE,
  • Gold

Material gasket

  • PTFE

Fill fluid

  • Silicone oil, Inert oil,
  • Vegetable oil,
  • High/ low temperature oil

Material housing

  • Aluminium,
  • 316L

Communication

  • 4...20 mA HART
  • PROFINET with Ethernet-APL
  • Bluetooth® wireless
  • technology and app (optional)

Certificates / Approvals

  • ATEX, CSA, IECEx, JPN Ex, INMETRO, NEPSI, UK Ex

Safety approvals

  • SIL

Design approvals

  • EN10204-3.1,
  • NACE MR0103,
  • NACE MR0175

Marine approvals

  • Marine approvals

Drinking water approvals

  • NSF

Specialities

  • Heartbeat Technology,
  • Bluetooth® Operation and maintenance SmartBlue App,
  • LED Module,
  • RFID TAG for easy identification,
  • Plug and play functionalities,
  • HistoROM
show more