Heavy duty (old thermometer line) T13
Brand: Endress+HauserHeavy duty (old thermometer line) T13
Brand: Endress+Hauser
Extended Specs at a Glance
Max. process pressure (static):
at 20 °C: 100 bar (1.450 psi)
depends on configuration
depends on configuration
Accuracy:
class A acc. to IEC 60751
class B acc. to IEC 60751
class B acc. to IEC 60751
Max. immersion length on request:
up to 22.5" (571 mm)
others on request
others on request
Response time:
depending on configuration
63% rt = 20 s
63% rt = 20 s
Operating temperature range:
PT100 WW:
-200 °C … 600 °C
(-328 °F … 1.112 °F)
PT100 TF:
-50 °C … 200 °C
(-58 °F … 392 °F)
-200 °C … 600 °C
(-328 °F … 1.112 °F)
PT100 TF:
-50 °C … 200 °C
(-58 °F … 392 °F)
Field of application
The robust device with a barstock thermowell is a complete unit ready for use with various housing transmitter heads to meet the application and space requirements. The unit offers enhanced measurement accuracy and reliability. A variety of process connections, dimensions, and materials (like 316L SS and Hastelloy C276) offer flexible application possibilities.
Features
- High flexibility due to modular assembly with standard terminal heads and customized immersion length
- One Source shopping for temperature measurement solutions. World class transmitter with integrated sensor offering for heavy process industry applications. Remove and install straight out of the box!
- Fast response time with reduced/tapered tip form
- Head transmitter with easy selection: Analog output 4 to 20 mA, HART®, PROFIBUS® PA or FOUNDATION Fieldbus™
Specifications
Brand
Industry
- Chemical
- Power & Energy
- Mining, Minerals & Metals
- Oil & Gas
- Other industries
Thermometer
Characteristic / Application
- Explosions Proof US style
- modular temperature assembly
- for heavy duty applications
- threaded process connection
- with extension
- incl. thermowell
Thermowell / protection tube
- bar stock (drilled)
Insert / probe
- mineral insulated (MI), flexible
- PTFE-insulated, rigid
Outer diameter protection tube / Insert
- 3/4'' (19,05 mm)
- 1'' (25,40 mm)
- 17/16'' (26,99 mm)
- 0,63'' ...1,31'' (15,88 mm ...33,40 mm)
Max. immersion length on request
- up to 22.5" (571 mm)
- others on request
Material protection tube/ thermowell
- 316/ 316L
- A105
- others on request
Optional coating
- available on request
Process connection
- male thread:
- NPT1/2"
- NPT3/4"
- NPT1"
- weld in version
- socket weld version
Tip shape
- stepped
- tapered
Surface roughness Ra
- 32 μin. (0.80 μm)
Operating temperature range
- PT100 WW:
- -200 °C … 600 °C
- (-328 °F … 1.112 °F)
- PT100 TF:
- -50 °C … 200 °C
- (-58 °F … 392 °F)
Max. process pressure (static)
- at 20 °C: 100 bar (1.450 psi)
- depends on configuration
Accuracy
- class A acc. to IEC 60751
- class B acc. to IEC 60751
Response time
- depending on configuration
- 63% rt = 20 s
Integration head transmitter
- yes (4 … 20 mA; HART; PROFIBUS PA; FOUNDATION FIELDBUS)
Ex - approvals
- FM XP DIP
- FM XP NI DIP
- CSA XP DIP
- CSA XP NI DIP
- FM/CSA XP DIP
- FM/CSA XP NI DIP
- CSA C/US XP, DIP
- cCSAus General Purpose
Certification
- SIL (transmitter only)