Specs at a Glance

Measuring method:
TOC determination by UV digestion and measurement of the differential conductivity
Process pressure:
max. 0.5 bar (7.25 psi)
Measuring range:
0.5 to 1 000 μg/l (ppb)
Process temperature:
< 50 °C (122 °F)

Field of application

TOC Analyzer CA79
Differential conductivity + UV digestion

Specifications

Analyser

Characteristic

  • Total carbon (TOC) analyzer for trace levels

Measuring method

  • TOC determination by UV digestion and measurement of the differential conductivity

Size

  • Housing:
  • 500 x 290 x 200 mm
  • 19.68 x 11.41 x 7.87 in

Design

  • Stainless steel housing;
  • IP 42 (standard), IP54 (optional)

Process temperature

  • < 50 °C (122 °F)

Ambient temperature

  • 10 to 45 °C (50 to 113 °F)

Process pressure

  • max. 0.5 bar (7.25 psi)

Sample flow rate

  • Min. 5 ml/min (0.17 fl.oz/min)

Consistency of the sample

  • max. conductivity 2 μS/cm, particle free

Specials

  • The operating concept and data storage meet the requirements of 21 CFR, Part 11
  • UV reactor with continuous function monitoring

Application

  • Determination of total carbon in ultrapure water applications, e.g. in the lifescience industry, that meet the following conditions:
  • Conductivity < 2 μS/cm
  • pH range: neutral

Power supply

  • 100/240 V AC, 47 - 63 Hz

Output / communication

  • 0/4 to 20 mA, galvanically isolated

Input

  • 1 to optional 3 measuring channels
  • Optional control input 24 V (for 1 channel instruments)

Measuring range

  • 0.5 to 1 000 μg/l (ppb)
show more