Thermometer Industrial temperature thermometer TM111

Brand: Endress+Hauser

Thermometer Industrial temperature thermometer TM111

Brand: Endress+Hauser

Rust Automation & Controls, Endress+Hauser's exclusive USA representative in:
 CA CO HI MT   Parts of ID ND NV   UT WY

Extended Extended

Specs at a Glance

Max. process pressure (static):
depending on the configuration
Accuracy:
Class AA acc. to IEC 60751
Class A acc. to IEC 60751
Class B acc. to IEC 60751
Class special or standard acc. to ASTM E230
Class 1 or 2 acc. to IEC 60584-2
Max. immersion length on request:
up to 4.500,0 mm (177'')
Response time:
t90 starting at < 1,5 s QuickSens
depending on configuration
Operating temperature range:
PT100 TF StrongSens:
-50 °C ...500 °C
(-58 °F ...932 °F)
PT100 QuickSensTF:
-50 °C …200 °C
(-58 °F …392 °F)
PT100 WW:
-200 °C ...600 °C
(-328 °F ...1.112 °F)
PT100 TF:
-50 °C ...400 °C
(-58 °F ...752 °F)
Typ K:
max. 1.100 °C
(max. 2.012 °F)
Typ J:
max. 800 °C
(max. 1.472 °F)
Typ N:
max. 1.100 °C
(max. 2.012 °F)

Field of application

Modular thermometer with resistance
sensor or thermocouple w/o thermowell
for direct process contact.
Metrical version with non-exchangeable
insert for demanding applications in a
high variety of industries.
Measuring range: -200...+1100oC/
-328...+2012oF
(depending on the configuration).
Consider additional marine certificate
requirements on application side.

Specifications

Target Customer

  • Chemical industry
  • Power & Energy industry
  • Oil & Gas industry

Product Family

  • iTHERM ModuLine

Thermometer

Characteristic / Application

  • metric style
  • universal range of application
  • suitable for hazardous areas
  • can be used with StrongSens, QuickSens insert
  • direct process contact

Thermowell / protection tube

  • without, direct process contact

Insert / probe

  • mineral insulated (MI), flexible

Outer diameter protection tube / Insert

  • Insert:
  • 3,0 mm
  • 6,0 mm

Max. immersion length on request

  • up to 4.500,0 mm (177'')

Material protection tube/ thermowell

  • Insert Material:
  • 316L (1.4404)
  • Alloy 600 (2.4816)
  • Pyrosil

Process connection

  • Thread:
  • G1/4, G1/2"
  • NPT1/2", NPT3/4"
  • M18x1.5, M20x1.5
  • Cap-nut:
  • G1/2", G3/4"
  • Compression fitting, also spring load:
  • NPT1/2", G1/2"
  • Weld-in adapter cylindrical or spherical

Tip shape

  • straight

Surface roughness Ra

  • < 1.6 μm (63.00 μin)

Operating temperature range

  • PT100 TF StrongSens:
  • -50 °C ...500 °C
  • (-58 °F ...932 °F)
  • PT100 QuickSensTF:
  • -50 °C …200 °C
  • (-58 °F …392 °F)
  • PT100 WW:
  • -200 °C ...600 °C
  • (-328 °F ...1.112 °F)
  • PT100 TF:
  • -50 °C ...400 °C
  • (-58 °F ...752 °F)
  • Typ K:
  • max. 1.100 °C
  • (max. 2.012 °F)
  • Typ J:
  • max. 800 °C
  • (max. 1.472 °F)
  • Typ N:

Max. process pressure (static)

  • depending on the configuration

Accuracy

  • Class AA acc. to IEC 60751
  • Class A acc. to IEC 60751
  • Class B acc. to IEC 60751
  • Class special or standard acc. to ASTM E230
  • Class 1 or 2 acc. to IEC 60584-2

Response time

  • t90 starting at < 1,5 s QuickSens
  • depending on configuration

Integration head transmitter

  • yes (4 … 20 mA; HART; PROFIBUS PA; FOUNDATION FIELDBUS)

Ex - approvals

  • ATEX
  • ATEX IECEx
  • NEPSI
  • IECEx
  • EAC Ex
  • CSA C/US
  • INMETRO

Certification

  • SIL (Transmitter)
show more