Memosens CPS92E
Brand: Endress+HauserMemosens CPS92E
Brand: Endress+Hauser
Extended Specs at a Glance
Process temperature:
0 to 110 °C (32 to 230 °F)
Measuring range:
–1 500 to 1 500 mV
Process pressure:
0.8 to 14 bar (11.6 to 203 psi) (absolute)
Field of application
Memosens CPS92E with platinum cap is specialized on media with a high content of suspended solids such as dispersions, precipitations or emulsions. The open aperture prevents fouling and ensures the most reliable measurement. Thanks to Memosens 2.0 digital technology, CPS92E combines maximum process integrity with simple operation. It resists moisture, enables lab calibration and offers extended storage of calibration and process data providing the perfect basis for predictive maintenance.
Chemical processes
Pulp and paper industry
Contaminated media and suspended solids- Dispersions- Precipitations- Emulsions
With ATEX, IECEx, CSA C/US, NEPSI, Japan, INMETRO approvals for use in hazardous areas, zones 0, 1 and 2.
Features
- Memosens 2.0 offers extended storage of calibration and process data, enabling better trend identification and providing a future-proof basis for predictive maintenance and enhanced IIoT services.
- Open aperture for application in heavily soiled media: Junction cannot be blocked by fibres or suspended solids.
- Not affected by fluctuations in pressure and temperature: Thanks to open aperture no diffusion potential can build up at the junction.
- Low maintenance and long lifetime due to stabilized gel filling
- Maximum process safety through non-contact inductive signal transmission
- Reduced operating costs thanks to minimized process downtime.
Specifications
Brand
Industry
- Chemical
- Pulp & Paper (incl. Printing)
- Other industries
Product Family
- Memosens
ORP / Redox
Application
- • Chemical processes
- • Pulp and paper industry
- • Contaminated media:
- • Solids
- • Emulsions
- • Precipitation reactions
- • Dispersions
Installation
- Installation
Characteristic
- Digital ORP electrodes with open aperture for contaminated media and integrated temperature sensor
Measuring range
- –1 500 to 1 500 mV
Measuring principle
- Gel compact electrode with open aperture and measuring part as platinum cap
Design
- All shaft lengths with temperature sensor
- Advanced gel technology
Material
- Sensor shaft: Glass to suit process
- ORP measuring element: Platinum
- Metal lead: Ag/AgCl
- O-ring: FKM
- Process coupling: PPS fiber-glass reinforced
- Nameplate: Ceramic metal oxide
Dimension
- Diameter: 12 mm (0.47 inch)
- Shaft lengths: 120, 225 and 360 mm
- (4.72, 8.86 and 14.17 inch)
Process temperature
- 0 to 110 °C (32 to 230 °F)
Process pressure
- 0.8 to 14 bar (11.6 to 203 psi) (absolute)
Temperature sensor
- NTC 30K
Ex certification
- With ATEX, IECEx, CSA C/US, NEPSI, Japan Ex and INMETRO approvals for use in
- hazardous areas Zone 0, Zone 1 and Zone 2.
Connection
- Inductive, contactless connection head with Memosens 2.0 technology
Ingress protection
- IP68
- Chemical Processes
- Pulp and Paper Industry
- Contaminated Media:
- Solids
- Emulsions
- Precipitation Reactions
- Dispersions