Thermometer iTHERM CompactLine TM311
Brand: Endress+HauserThermometer iTHERM CompactLine TM311
Brand: Endress+Hauser
Fundamental
Specs at a Glance
Max. immersion length on request:
up to 600,00 mm (23,62'')
Operating temperature range:
PT 100:
-50 °C ...200 °C
(-58 °F ...392 °F)
-50 °C ...200 °C
(-58 °F ...392 °F)
Max. process pressure (static):
at 20 °C: 50 bar (725 psi)
Response time:
t50 = 1 s
t90 = 1.5 s
t90 = 1.5 s
Accuracy:
class A acc. to IEC 60751
Field of application
The iTHERM CompactLine TM311 is a thermometer that provides accurate and fast process temperature measurements in hygienic and aseptic applications. It is available with a common Pt100 RTD or a Pt100 RTD with TipSens technology reaching response times as low as 1.5 s (t90) for optimum process control. The sensor is very accurate even with short immersion lengths.
Features
- Version with iTHERM TipSens offers best-in-class response times with a T90 time of 1.5 seconds
- Version with iTHERM TipSens ensures a highly accurate measurement for short immersion lengths for example in small pipelines in the food and beverages industry
- Fast installation and easy commissioning thanks to compact dimensions for installation in tight spaces
- 4…20mA and IO-Link in one device. Perfect to save costs for stock
Specifications
Brand
Field of Application
- Convenience food
- Bodycare, soap, detergent and healthcare
- Dairy and dairy products
- Bioreactors
Industry
- Chemical
- Water, Wastewater & Waste
- Food & Beverage
- Power & Energy
- Other industries
- Life Science
Target Customer
- Life sciences industry
- Food & beverage industry
- Waste water treatment plants
- Meat and fish producer
- Power & Energy industry
Product Family
- iTHERM CompactLine
Thermometer
Characteristic / Application
- metric and imperial style
- compact temperature probe
- fast response time
- high accuracy
- standard and hygienic design
- hygienic process connection
- threaded process connection
Thermowell / protection tube
- without
- incl. thermowell
- incl. Tee-/elbow thermowell
Insert / probe
- pipe version, isolated wires, not flexible
Outer diameter protection tube / Insert
- 3,0 mm (0,12'')
- 6,0 mm (0,24'')
Max. immersion length on request
- up to 600,00 mm (23,62'')
Material protection tube/ thermowell
- 1.4435 Delta-Ferrit < 1%
- 316L
Process connection
- compression fitting
- Thread
- weld in adapter
- clamp connections acc. to ISO2852
- screwed pipe joint acc. to DIN 11851
- aseptic screwed pipe joint acc. to DIN 11864-1
- metallic sealing system
- thread acc. to ISO 228 for liquiphant adaptor
- Varivent
- SMS 1147
- Tee-/elbow thermowell
Tip shape
- straight
- reduced
Surface roughness Ra
- 0,76 μm (29,92 μin)
- 0,38 μm (14,96 μin)
- 0,38 μm (14,96 μin) electropolished
Operating temperature range
- PT 100:
- -50 °C ...200 °C
- (-58 °F ...392 °F)
Max. process pressure (static)
- at 20 °C: 50 bar (725 psi)
Accuracy
- class A acc. to IEC 60751
Response time
- t50 = 1 s
- t90 = 1.5 s
Integration head transmitter
- no (4…20mA Signal, IO-Link)
Certification
- 3A, EHEDG, ASME BPE, FDA, TSE
- 1935/2004, 2023/2006, 10/2011
- DNVGL, BV
- CSA C/US General Purpose
- RCM