Oxymax COS61

Brand: Endress+Hauser

Oxymax COS61

Brand: Endress+Hauser

Rust Automation & Controls, Endress+Hauser's exclusive USA representative in:
 CA CO HI MT   Parts of ID ND NV   UT WY

Specs at a Glance

Process temperature:
-5°C ... 60°C
( 23 - 140°F)
Process pressure:
max. 10bar
(145psi)
Measuring range:
0mg/l ... 20mg/l
0%SAT ... 200%SAT
0 ... 400hPa

Field of application

Oxymax COS61 is a high-performance oxygen sensor that offers fast, accurate and drift-free measurement. It supports your process with low maintenance, high availability and easy handling. The sensor's long-term stable fluorescence layer is exclusively oxygen-selective (interference-free), ensuring consistently reliable measurement.

Oxymax COS61 measures the dissolved oxygen concentration in:

  • Wastewater treatment plants:- O2 control in the aeration basin- Process water treatment and monitoring
  • Water works:- Status monitoring of drinking water (oxygen enrichment, corrosion protection, etc.)- Water quality monitoring in rivers, lakes or seas
  • Utilities of all industries:- O2 control in the biological treatment- Process water treatment and monitoring
  • Fish farming:- O2 control for optimum growth conditions

Features

  • Minimum maintenance, maximum availability
  • Fast, drift-free measurement for aeration control and process monitoring
  • Long-term stability for increased process safety
  • High-performing in all aeration processes (SBR, Anamox, etc.)
  • Chemical-free: no electrolyte handling
  • Easy measuring point changeover to optical technology compatible with COS31 and COS41

Specifications

Industry

  • Chemical
  • Water, Wastewater & Waste
  • Food & Beverage
  • Power & Energy
  • Mining, Minerals & Metals
  • Oil & Gas
  • Pulp & Paper (incl. Printing)
  • Other industries
  • Life Science

Product Family

  • Oxymax

Oxygen

Application

  • Aeration tank, river monitoring, water treatment, fish farming.

Characteristic

  • Digital, optical (luminescence) dissolved oxygen measurement.
  • No flow needed - measurement possible in still water.

Measuring range

  • 0mg/l ... 20mg/l
  • 0%SAT ... 200%SAT
  • 0 ... 400hPa

Measuring principle

  • Oxygen-sensitive molecules (marker) are integrated in an optical active layer (fluorescence layer). The fluorescence layer surface is in contact with the medium. The sensor optics are directed at the back of the fluorescence layer. The sensor optics transmit green light pulses to the fluorescence layer. The markers respond (fluoresce) with red light pulses. The duration and intensity of the response signals depend directly on the oxygen contents or partial pressure.

Design

  • - Calibration data saved in sensor.
  • High degree of EMC protection.

Material

  • Sensor body : stainless steel 1.4571Membrane cap : POM

Dimension

  • Diameter : 40mm
  • (1.56inch)Length : 220mm
  • (8.58inch)

Process temperature

  • -5°C ... 60°C
  • ( 23 - 140°F)

Process pressure

  • max. 10bar
  • (145psi)

Temperature sensor

  • NTC temperature sensor, 0 to 50°C
  • (32 - 122°F)

Connection

  • Process connection: G1" thread
  • Cable connection : fixed cable or TOP68 plug-in-head.
show more