pH sensor Memosens CPS11E
Brand: Endress+HauserpH sensor Memosens CPS11E
Brand: Endress+Hauser
Extended Specs at a Glance
Process pressure:
Applications A and B:
0.8 to 17 bar (11.6 to 246.5 psi) absolute
Application F:
0.8 to 7 bar (11.6 to 101.5 psi) absolute
0.8 to 17 bar (11.6 to 246.5 psi) absolute
Application F:
0.8 to 7 bar (11.6 to 101.5 psi) absolute
Process temperature:
Application A: –15 to 80 °C (5 to 176 °F)
Application B: 0 to 135 °C (32 to 275 °F)
Application F: 0 to 70 °C (32 to 158 °F)
Application B: 0 to 135 °C (32 to 275 °F)
Application F: 0 to 70 °C (32 to 158 °F)
Measurement range:
Application A
• pH: 1 to 12
Application B
• pH: 0 to 14
Application F
• pH: 0 to 10
• pH: 1 to 12
Application B
• pH: 0 to 14
Application F
• pH: 0 to 10
Field of application
The Memosens CPS11E, a digital pH sensor, guarantees process stability and provides reliable measured values, even in harsh pH environments or in risk zones. Thanks to Memosens 2.0 digital technology, this sensor is moisture resistant and supports laboratory calibration, which increases process safety and reduces operational complexity. With extended storage of calibration and process data, it provides a solid basis for predictive maintenance. This sensor is designed for durability and minimal maintenance, ensuring a long service life.
Features
- Robust electrode: Special design prevents poisoning, with a dirt-repellent PTFE junction guarding against medium contamination.
- Durable materials: Process glass handles full pH range, tolerates up to 17 bar (246.5 psi) pressure; Optional F-glass suits hydrofluoric acid, like in semiconductor baths.
- Fast sensor replacement reduces process downtime, extends sensor life and minimizes operating costs.
- Increased reliability in low conductivity applications: An improved optional salt storage ensures reliable measurements in scenarios such as boiler feed water.
- Memosens 2.0 technology ensures data storage for better trend analysis, predictive maintenance, and advanced IIoT functions.
Specifications
Brand
Field of Application
- Battery components production
- Water monitoring, production and distribution
- Industrial gases
Target Customer
- Waste water treatment plants
- Pulp & Paper industry
- Refineries
Product Family
- Memosens
pH
Application
- Long-term monitoring and limit control in processes with stable process conditions
- Water and wastewater treatment
Characteristic
- Digital pH electrode for standard applications in process and environmental technology with dirt-repellent PTFE ring junction and built-in temperature sensor
Measurement range
- Application A
- • pH: 1 to 12
- Application B
- • pH: 0 to 14
- Application F
- • pH: 0 to 10
Measuring principle
- Gel compact electrode with PTFE ring junction
Design
- All shaft lengths with temperature sensor
- Advanced gel technology
Material
- Sensor shaft: Glass to suit process
- pH membrane glass: Type A, B, F
- Metal lead: Ag/AgCl
- Aperture: Ring-shaped PTFE junction, sterilizable
- O-ring: FKM
- Process coupling: PPS fibre-glass reinforced
- Nameplate: ceramic metal oxide
Dimension
- Diameter: 12 mm (0.47 inch)
- Shaft length: 120, 225, 360 and 425 mm
- (4.72, 8.86, 14.17 and 16.73 inch)
Process temperature
- Application A: –15 to 80 °C (5 to 176 °F)
- Application B: 0 to 135 °C (32 to 275 °F)
- Application F: 0 to 70 °C (32 to 158 °F)
Process pressure
- Applications A and B:
- 0.8 to 17 bar (11.6 to 246.5 psi) absolute
- Application F:
- 0.8 to 7 bar (11.6 to 101.5 psi) absolute
Temperature sensor
- NTC 30K
Ex certification
- With ATEX, IECEx, CSA C/US, NEPSI, Japan Ex and INMETRO approvals for use in
- hazardous areas Zone 0, Zone 1 and Zone 2
Connection
- Inductive, digital connection head with Memosens 2.0 technology
Ingress protection
- IP68
Additional certifications
- Additional certifications
- Chemical Industry
- Pulp & Paper Industry
- Power Plants (e.g. Flue Gas Washers, Boiler Feed Water)
- Incineration Plants
- Water & Wastewater Treatment