MS01 pH Sensor Cube
Brand: Burkert Fluid Control SystemsThe device is a pH measurement sensor. It is used within the Online Analysis System Type 8905 by being plugged into a spare fluidic backplane slot.
The pH value is the most common parameter in water analysis. The pH sensor cube contains the ISFET measuring cell, based on Microelectromechanical systems technology (MEMS) to measure the pH value. An Ag/AgCl 3-mol KCl/l electrode is used as reference.
The electrical and fluidic connections are made via the connection panel of the system. The sensor cube is communicating via buS, so the recognition at the Online Analysis System is fully automatic. When plugging into a system you will find the sensor in the list of buS members for further customized adjustments.
Features
- Fully compatible with buS systems and a wide range of further analysis sensor cubes
- Sensor: MEMS ISFET technology
- Modular sensor cube for hot swap (exchange during operation)
- Minimal sample water flow needed
Specifications
Housing Material
- PPE+PS
Seal Material
- EPDM
Lever
- Zamak, Painted
Measuring Range
- pH 4 to pH 9
- 0° to 50° C (32° to 122° F)
Measuring Deviation
- ± 0.1 pH
Measuring Range Resolution
- 0.02 pH
- 0.01° C (0.018° F)
Linearity & Repeatability
- ± pH 0.05
Response Time
- < 10 s
Environmental Protection
- IP20
- IP65
Operating Voltage
- 24 Vdc
Power Consumption
- 0.8 VA
Media
- Water Without Particles: Drinking Water, Industrial Water
- For Cl < 0.2 ppm, use anti-biofouling external reference electrode
Medium Temperature
- Sample Fluid: 3° to 40° C (37° to 104° F)
Approvals & Certifications
- CE
- EU Directives
- UL (C,US)
Flow Rate
- Sample Water: > 6 L/hr
Process Connection
- Via pinch valve in the fluidic backplane of Type 8905
Temperature Range
- 0° to 50° C (32° to 122° F)
Ambient Temperature
- 0° to 40° C (32° to 104° F)
Humidity
- < 90 % RH, without Condensation
Communication
- Via büS (Bürkert system bus, CANopen-based) or CANopen
Electrical Connection
- Spring Contacts
- Indoor & Outdoor
- Protect the Device Against Electromagnetic Interference, Ultraviolet Rays
- When Installed Outdoors, Against the Effects of Climatic Conditions