Specs at a Glance

Measured Variables:
Concentration
Cell Pressure
Cell Temperature
Measuring principle:
Quenched Fluorescence
Hazardous area approvals:
ATEX / IECEx - Zone 2
CSA Class I, Division 2

Field of application

The OXY5500 oxygen analyzer is a compact, stand-alone, single channel analyzer that uses quenched fluorescence (QF) technology to reliably measure oxygen in natural gas and gas processing applications. A sensor probe is inserted into the process stream and is connected to the controller by an optical fiber. Featuring an LCD display and built-in data logger, the OXY5500 also has full sample conditioning systems available as well as software for a PC interface.

Natural gas quality measurements in production, storage, transportation, and distribution

Monitoring in natural gas, upstream of production and gathering sites

Measurable range from ppm levels to 0-20% O2

Trace measurements as low as 0-10 ppmv oxygen in hydrogen for electrolyzer applications

Ideal for CCUS applications since optical oxygen measurement is unaffected by CO2 or H2S in the process

Features

  • Easy to navigate display and menu
  • Small optical sensor with no membrane or consumable chemicals
  • Optical measurement with fast, continuous response
  • Excellent long-term stability
  • Not affected by H2S
  • No moving parts, simple operation
  • Data logging for 30 days

Specifications

Industry

  • Oil & Gas

O2

Product Headline

  • Compact, single channel oxygen (O2) analyzer with an LCD display and built-in data logger. The sensor measures oxygen using quenched fluorescence (QF) technology, which is inserted into the process stream and is connected to the controller by an optical fiber.

Channels

  • 1

Analyte and Measurement ranges

  • O2 (Oxygen):
  • 0 to 100 ppmv (minimum)
  • 0 to 20% volume (maximum)

Measured Variables

  • Concentration
  • Cell Pressure
  • Cell Temperature

Ambient Temperature range

  • -20 to 50°C (-4 to 122°F)

Operating Pressure range

  • Inlet Pressure: 140 to 275 kPa (20 to 40 psig)
  • Sample Cell: 800 to 1200 mbara

Analyzer Wetted materials

  • 304 & 316 stainless steel
  • FKM O-Rings

Power supply

  • 108 to 253 VAC, 50/60 Hz; 5.3 Watts at 120VAC; 6.6 Watts at 240VAC
  • or
  • 9 to 30 VDC (CSA), 18 to 30 VDC (IECEx/ATEX); 4.7 Watts at 24VDC

Communication

  • Analog Output: Two 4-20 mA outputs
  • Analog Input: One 4-20 mA input (sample pressure)
  • Fieldbus: RS-232C, RS-485, & Ethernet 10/100 with Modbus
  • Output Relays: Qty 2, 250 mA max load (Concentration and Fault Alarms)
  • USB 2.0 Works With Service Software Only
  • 4 GB Internal Memory with Internal Data Logging

Housing materials

  • Electronics: 304 or 316L stainless steel

Hazardous area approvals

  • ATEX / IECEx - Zone 2
  • CSA Class I, Division 2

Degree of protection

  • IP64, Type 4X

Product safety

  • CE
  • RCM
  • FCC
show more