Condumax CLS13
Brand: Endress+HauserCondumax CLS13
Brand: Endress+HauserSpecs at a Glance
Process pressure:
max. 40 bar (580 psi)
Process temperature:
max. 250 °C (482 °F)
Measuring range:
k=0,01: 0-20,0 µS/cm
k=0,1: 0-200,0 µS/cm
k=0,1: 0-200,0 µS/cm
Field of application
Condumax CLS13 is a robust conductivity sensor for steam/water cycles in power plants. It is optimized for low conductivities (even under very high temperatures and pressures) to protect your processes from deposits and corrosion. Low maintenance and simple to commission (large connection compartment), this is the ideal sensor for tough environments.
Condumax CLS13 measures conductivity in the steam/water cycle in the power & energy industry:
- Condensate monitoring
- Boiler feedwater monitoring
- Boiler blowdown control
- The sensor can be used in all low conductivity applications with very high temperatures and pressures.
- With its international explosion protection approvals, such as ATEX and FM, the sensor is suitable for hazardous area use.
Features
- Reliable and accurate measuring values at low conductivities
- Optimized for high pressures and temperatures up to 250°C (482°F)
- Easy cleaning thanks to stainless steel body and removable outer electrode
- Robust design for high durability
- Quality certificate stating the individual cell constant
Specifications
Brand
Industry
- Chemical
- Power & Energy
- Mining, Minerals & Metals
- Other industries
Product Family
- Condumax
Conductivity
Application
- Process, power plant
Characteristic
- 2-electrode conductivity cell for high temperature applications.
Measuring range
- k=0,01: 0-20,0 µS/cm
- k=0,1: 0-200,0 µS/cm
Measuring principle
- Coaxially arranged electrodes for high temperature applications
Design
- 2-electrode conductivity cell with coaxially arranged electrodes
- Connection head and cooling body made of casting aluminium with fixed cable connection PG16
Material
- Electrode: 1.4571
- Sealing: Kalrez, ceramic
Dimension
- Electrode diameter: 25 mm (0.984 inch)
- Immersion depth: 50 to 115 mm (1.97 to 4.48 inch)
- Weight: 3kg
Process temperature
- max. 250 °C (482 °F)
Process pressure
- max. 40 bar (580 psi)
Temperature sensor
- Integrated Pt100
Ex certification
- ATEX
Connection
- Process connection: G1"
Ingress protection
- IP67
Additional certifications
- Calibration certificate with cellconstant