Fluke 287 True-RMS Electronics Logging Multimeter

Brand: Fluke Corporation

The Fluke 287 True-rms Electronics Logging Multimeter with TrendCapture quickly documents design performance and graphically displays what happened.

Its unique logging and graphing capabilities mean you no longer need to download logged readings to a PC to detect a trend.

The Fluke 287 packs more accuracy and convenience into a handheld multimeter than ever before.
  • Real time clock for automatic time stamping of saved readings.
  • True-RMS AC voltage and current for accurate measurements on complex signals or non-linear loads. AC bandwidth specified to 100 kHz.
  • Measure up to 10 A (20 A for 30 seconds).
  • 100 mF capacitance range.
  • Temperature function.
  • Relative mode to remove test lead resistance from low ohms or capacitance measurements. • Peak capture to record transients as fast as 250 μs.
  • Premium test leads and alligator clips included.
  • Amp jack plugs included.
  • Optional FlukeView forms enable you to document, store and analyze individual readings or a series of measurements, then convert them into professional-looking documents.
  • Optional magnetic hanger allows you to hang the meter for easy viewing while freeing your hands to focus on the job
Features
  • Now compatible with Fluke Connect mobile app and all Fluke FC enabled test tools with optional ir3000 FC infrared connector (sold separately).
  • Let your team see what you see in an instant with ShareLive™ video call (requires Fluke Connect mobile app and ir3000 FC wireless connector).
  • TrendCapture quickly graphically displays logged data session to quickly determine whether anomalies may have occurred.
  • Limited lifetime warranty.

Specifications

DC Volts

  • Range / resolution: 50.000 mV, 500.00 mV, 5.0000 V, 50.000 V, 500.00 V, 1000.0 V
  • Basic accuracy: 0.025 %

AC Volts

  • Range / resolution: 50.000 mV, 500.00 mV, 5.0000 V, 50.000 V, 500.00 V, 1000.0 V
  • Basic accuracy: 0.4 % (True-RMS)

DC Current

  • Range / resolution: 500.00 μA, 5000.0 μA, 50.000 mA, 400.00 mA, 5.0000 A, 10.000 A
  • Basic accuracy: 0.05 %

AC Current

  • Range / resolution: 500.00 μA, 5000.0 μA, 50.000 mA, 400.00 mA, 5.0000 A, 10.000 A
  • Basic accuracy: 0.6 % (True-RMS)

Temperature

  • Range / resolution: -200.0° C to 1350.0° C
  • Basic accuracy: 1.0 %

Resistance

  • Range / resolution: 500.00 Ω, 5.0000 kΩ, 50.000 kΩ, 500.00 kΩ, 5.0000 MΩ, 50.00 MΩ, 500.0 MΩ
  • Basic accuracy: 0.05 %

Capacitance

  • Range / resolution: 1.000 n, 10.00 nF, 100.0 nF, 1.000 μF, 10.00 μF, 100.0 μF, 1000 μF, 10.00 mF, 100 mF
  • Basic accuracy: 1.0 %

Frequency

  • Range / resolution: 99.999 Hz, 999.99 Hz, 9.9999 kHz, 99.999 kHz, 999.99 kHz
  • Basic accuracy: 0.005 %

Operating Temperature

  • -20° C to 55° C

Vibration

  • Random vibration per MIL-PRF-28800F Class 2
show more